W. M. Szymczyk
Soltan Institute for Nuclear Studies, 05-400 Świerk/Otwock, Poland
The paper describes basic steps of the automatic element identification algorithm implemented in the
"KSAF" and "SRDE" Energy Dispersive X-ray Fluorescence (EDXRF) analytical systems manufactured in
the Soltan Institute for Nuclear Studies. The aim of the algorithm was to establish - without any
operator's intervention - a list of elements present in the analyzed sample, provided that a list
of peaks is already available {such a list is produced in KSAF and SRDE systems by a spectrum
analysis algorithm described elsewhere). Each, analytical peak is assigned a set of line/element
identification proposals (ID's) based on built-in libraries of characteristic X-ray energies; all the
ID's form a list of candidate elements for the analyzed sample. According to the observed sequence
of lines the elements on the list are divided into three categories: elements DISREGARDED (a higher
intensity line missing), elements PROBABLE (a sequence of lines detected), elements POSSIBLE (those
which can neither be disregarded nor qualified as PROBABLE). Due to relatively poor energy resolution
of EDXRF systems (in comparison with e.g. wavelength dispersive XRF systems) the automatically produced
results may in many cases remain inconclusive. Therefore it was found that the implemented and
described herein algorithm is best used as a tool helping to concentrate operator's attention on
those assignments "peak in spectrum <=> element in sample" which are doubtful or difficult for
automatic establishing.