A COMPARISON OF THREE MATERIALS USED IN ESR DOSIMETRY: L-alpha-ALANINE, DL-alpha-ALANINE AND STANDARD BONE POWDER. RESPONSE TO Co-60 GAMMA RADIATION
(PORÓWNANIE ODPOWIEDZI TRZECH MATERIAŁÓW UŻYWANYCH W EPR DOZYMETRII: L-alfa-ALANINY, DL-alfa-ALANINY I STANDARDOWEGO PROSZKU KOSTNEGO NA PROMIENIOWANIE GAMMA Co-60)
Zofia Stuglik, Jaroslaw Sadlo
Three solid state materials: L-alpha-alanine, DL-alpha-alanine and standard bone powder were irradiated with gamma analyzed with ESR method.
It was stated that the G-value of paramagnetic centres in L-alpha-alanine is practically the same as in DL-alpha-alanine and about 50 times higher than in non-deproteinized bone powder.
The sensitivities of investigated materials are proportional to their G-values if double integrals of ESR signals are chosen as a measure of radiation effects.
When first derivatives of ESR absorption bands are used to the construction of dose-response curves (peak-to-peak method) the sensitivities of all investigated materials are comparable.
Trzy materiały: L-alfa-alaninę, DL-alfa-alaninę i standardowy proszek kostny napromieniano promieniowaniem gamma i badano metodą EPR. Stwierdzono, że wydajność radiacyjna centrów paramagnetycznych G, jest taka sama w L-alfa-alaninie jak w DL-alfa-alaninie i około 50 razy wyższa niż w proszku kostnym nie poddawanym deproteinizacji. Czułość badanych materiałów jest proporcjonalna do zmierzonych wartości G, gdy jako miarę efektów radiacyjnych stosuje się całkę podwójną z sygnału EPR. W sytuacji, gdy sygnałem analitycznym służącym do konstrukcji zależności dawka-efekt jest amplituda pierwszej pochodnej krzywej absorpcji mikrofalowej czułości wszystkich badanych materiałów są porównywalne.