APPLICATION OF MULTIVARIATE CALIBRATION FOR DETERMINATION OF CALCIUM AND IRON IN ASH WITH XRF ANALYSIS

Zastosowanie wielowymiarowej kalibracji w oznaczaniu wapnia i żelaza w popiołach metodą analizy fluorescencyjnej

Ewa Kowalska, Piotr Urbański


Multivariate calibration based on partial least square regression (PLS) was applied for determination of calcium and iron in the flying ashes from the brown coal, using low-resolution XRF analysis. The measurements were performed using an AF-20 XRF analyser with the Ar proportional counter and the Cd-109 radioisotope source. With the range of CaCO3 concentration from 20% to 60% the mean square error of crossvalidation was about 1.5% CaCO3 and for Fe2O3 of concentration ranging from 3% to 12% the RMSECV was lower then 0.3% Fe2O3. It was found that application of the PLS calibration considerably reduces matrix effect. Applied calibration method as well as used measuring device can be used also for determination of strontium, which is sometimes present in the flying ash samples.