DOSIMETRIC RESPONSE OF UNTINTED, COMMERCIALLY AVAILABLE CTA FOILS FOR 60Co GAMMA RAYS

OdpowiedŸ dozymetryczna nie barwionych folii z trioctanu celulozy (CTA) na promieniowanie gamma 60Co

Zofia Peimel-Stuglik


The usefulness of two kinds of untinted CTA foils: a Fuji CTR-125 dosimetric foil and a technical CTA-T foil, produced by "Zak³ady Chemiczne, Gorzów Wielkopolski" as support for light-sensitive layers of amateur photo-films, for 60Co gamma ray dosimetry was investigated. In spite of rather bad physical parameters of the technical foil (spread of foil thickness, high and different initial absorbance) the dosimetric response of both foils for 60Co gamma rays was similar. The CTA-T foil can be used for rutine dosimetry providing that dosimetric signals have to be calculated exactly as recommended by the ASTM standard, i.e. as the difference of absorbance of irradiated and (the same) non-irradiated foil. Any other approach may lead to high errors of dose evaluation. The last is true also for other CTA foils, especially after long shelf-life.